Question

I use Lecroy LT264 oscilloscope to trace lightning impulse test (1.2/50usec waveform). I have driver for labview and use edge triggered example. One impulse is not shown in every three impulse. What could be the reason ? Thanks for your interest.

enter image description here

enter image description here

Also you can download code here : http://forums.ni.com/t5/Instrument-Control-GPIB-Serial/LeCroy-LT264-Lightning-Impulse-Test-Signal-Loss/td-p/2510994

Was it helpful?

Solution

Ok, I solve the problem. The reason is trigger level. I had set 40 V or 80 V. However when I set trigger level 150V, program shows all impulse.

OTHER TIPS

Check to see how many samples you are recording. If the time it takes to collect samples is larger than the time spacing of the the trigger, then the hardware won't reset until all the samples are collected. This means the trigger won't get rearmed.

Try reducing the number of samples collected if you need to trigger on every impulse.

Also, seeing your code might help.